Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)Photo by Mackenzie Willborn (@mackenziew_EP)
Mitral AFL utilizing Bachmanns Bundle fibers as an epicardial connection - let’s get into it! Upon ablating an anterior mitral line, we extend our CL by 65ms. Further LA ablation proves no change to TCL - We ultimately terminate AFL at the septal RA/SVC junction by contact alone!
10
26
2
108
Mackenzie Willborn retweeted
LSPV gap and slow wave speed correlation? LSPV gap rapidly identified with Omnipolar Technology (OT) and isolated with #TactiCathSE. The wave speed map was analyzed and the slow zone (red-yellow) directly correlated with the remaining connection. @AbbottCardio #AbbottAF #EPeeps
Mackenzie Willborn retweeted
Timing in AF? Omnipolar Technology (OT) shows local wavefront activation independent of LAT. #OT and #HDGrid rapidly identified the leak in our right WACA. 🎯 look at that independent firing following isolation with #TactiCathSE 👀 #EPeeps #AbbottAF @AbbottCardio @RajeevJoshi400
1
17
73
Mackenzie Willborn retweeted
First VoXel AF case for this #EnSiteX user. The inherently linear geometry definitely assisted with navigation throughout this branchy chamber. 🧲 Look at that LSPV takeoff! 👀 #AbbottAF @AbbottCardio @mackenziew_EP
Mackenzie Willborn retweeted
Make your typical flutters eXciting with Omnipolar Technology (OT)! OT rapidly identified the narrow isthmus allowing this CTI-dependent flutter to persist. By using localized signal properties, OT vectors are independent of LAT annotation! #OnlyWithOT #EnSiteX #EPeeps
36
95
Mackenzie Willborn retweeted
With #EnSiteX and #OT we were able to see the muscle sleeve entering the LIPV — traditional bipolar EGMs showed a much broader connection. We’re seeing vein connections differently #OnlyWithOT. #EPeeps #TrueEGMs #HDGrid @AbbottCardio @OneBurnJuan @EPDenault
66
2
127
The open window mapping technique allowed us to create a very comprehensive map of this LLAP while in NSR with pre-excitation! I love being able to achieve this level of understanding of a pts arrhythmia while keeping them in comfortable sinus rhythm 🤗 @AbbottCardio #EPeeps
Mackenzie Willborn retweeted
By self-referencing sharp signals on #HDGrid, #LiveView allowed for beat-by-beat timing localization of the leak into the LIPV despite being in AF. Following white (early) with #TactiCathSE allowed for rapid isolation of the LIPV. @mackenziew_EP @AbbottCardio #JustGridIt #EPeeps
1
14
64
Color propagation for this interesting case 😊
6
Ablation at our termination site in the RA renders our Mitral AFL non-inducible 🤩 What an interesting case!
5
Right sided scar-related AFL mapped with #HDGrid 🤩 Surface EKG was consistent with Typical AFL, but this high density map gave us a clear image of what was truly propagating our patients reentrant circuit! @AbbottCardio
4
8
1
44
Critical isthmus of another Atypical Flutter looking GREAT with the help of some of my favorite Ensite Precision features - Sparkle map, a color contour adjustment, and Low V-ID adjustment to 0.70 mV. Beautiful map made with #HDGrid by @BurnThatSpot !!
1
7
2
33
More LiveView for confirming CTI ablation block! Here we have one beat paced from the CS & displayed from #HDGrid using #LiveView. We can confirm LiveView’s viability by using proximal electrodes of our duo-deca catheter in this case. LiveView map settings included 😊
2
1
19
#HDGrid giving us a beautiful target area for today’s LA AFL. Slow conduction right through the mitral isthmus 😍 @AbbottCardio
4
4
27
Using #LiveView with #HDGrid to confirm cavotricuspid isthmus block! Activation down the lateral wall can be appreciated in just a few beats - with none of the catheter movement or measuring required in the more standard differential pacing protocol! @AbbottCardio
1
20
Mackenzie Willborn retweeted
Utilise EnSite LiveView Dynamic Display to detect rotor in atrial fibrillation #AbbottEmp #HDGrid #LiveView #Epeeps @EricTSLim
1
38
6
80
Mackenzie Willborn retweeted
ILAM on the 👈🏼, AAFL on the 👉🏼. pt. with a history of AAFL presented to the lab in NSR. A CS paced ILAM map was collected to identify deceleration zones. AAFL was induced while collecting ant. data with #HDGrid. Can you see a correlation? @AbbottCardio @GarabelliPaul #EPeeps
1
10
2
55
Love seeing these fractionated, mid-diastolic signals on #HDGrid while mapping an interesting Atypical AFL. Here we have some spanning roughly 60% of the tachycardia cycle length! Points located in the white/red color band area. @AbbottCardio
2
9
3
55